Korean equipment and chemicals
[ Laboratory,Scientific Equipment ] (import)                                                                                   ÁÖ¹®Çϼ¼¿ä
Lubricating Greases Testing
 
Falex High Performance Multi-Specimen Test Machine

ASTM Specifications : D-2266, D-3702, D-4172
See also for more information :
Model 6 : Multi-Specimen Test Machine
 
High Performance Multi-Specimen Test Machine With Digital Instrumentation System
 
Features and specifications
Speed & Velocity:
 
Ten turn variable speed control dial. Seven way speed range from 10 to 300 RPM, 10 to 450 RPM, 25 to 900 RPM, 50 to 1800 RPM, 100 to 3600 RPM, 150 to 7200 RPM and 300 to 10,800 RPM, ¡¾0.5% accuracy*. With optional adapter, speeds as low as 0.01 are possible.
 
Load:
 
Applied through a pneumatic system. Dual range loading from 100 to 2000 lbs. And 50 to 1000 lbs. ¡¾2% accuracy*. With low load option from 5 to 60 lbs., ¡¾2% accuracy*.
 
Temperature Control:
 
Controlled system with digital proportional indicator/controller, high temperature limit cut-off, heater cup, special ceramic high temperature insulators and automatic temperature start-up. Temperature control from ambient to 870¡¾3¡ÆC. (1600¡¾5¡ÆF), optionally to 1093¡ÆC (2000¡ÆF). Specially designed water cooled test shaft. Optional refrigeration system cools reservoir fluid from ambient to -30¡ÆC (-22¡ÆF) for circulation through optional cooling chamber. When using Standard Multi-specimen adapters, temperature limit is 232 ¡ÆC (450¡ÆF).
 
Environments:
 
Wet or dry. Enclosed chamber for containing humidity or circulating special atmospheres and fluids.
 
Motion:
 
Unidirectional or optional oscillatory, both high and low speed. Low speed from 50 to 600 cycles per minute, 0¡Æ - 90¡Æ. High Speed:
 
 
 
Degrees of Oscillation
Cycles per Minute Ranges
 
0¡Æ - 20¡Æ
50 - 1500
21¡Æ - 30¡Æ
 50 - 1200
31¡Æ - 40¡Æ
50 - 1100
41¡Æ - 50¡Æ
50 - 900
51¡Æ - 70¡Æ
50 - 600
71¡Æ - 90¡Æ
50 - 500
 
 
Optional secondary controllable drive system, perpendicular to main test shaft for controlling speed of rollers when testing with infinitely variable roll/slide adapter for evaluating gear system materials.
 
*All accuracies are percent of full scale. Other accuracies available on special quote.
 
Torque:
 
Measured by load cell with digital panel meter and digital printer. Automatic high limit torque cut-off. Field calibration system. Optional low range for greater accuracy in measuring torque from 0 to 25 lbs-in ¡¾2% full scale accuracy.
 
Available Speed Ranges
 
(Customer to choose from available ranges)
 
Spindle Speed Maximum Available
 
Ratio
RPM
Torque (lbs-in)
1:6
 
1:4
 
1:2
 
1:1
 
2:1
 
4:1
 
6:1
10-300
 
10-450
 
25-900
 
50-1800
 
100-2600
 
150-7200
 
300-10800
450
 
420
 
210
 
105
 
52.5
 
26.26
 
13.1
 
 
Wear:
 
Continuous electronic wear-indicating system incorporating an LVDT and digital indicator. (Reads to 0.0001 inch).
 
Test Time:
 
Digital timer/controller for automatic start-up or shut-off.
 
Cycle Counter:
 
Panel mounted cycle counter with automatic cut-off. (999,999 cycles).
 
Motor:
 
3 HP, variable speed DC motor.
 
Standard High Performance Multi-Specimen Configurations:
 
Area Contact: Thrust Washer
Line Contact: Vane-on-Disk
Point Contact: Pin-on-Disk
 
Optional High Performance Multi-Specimen Configurations:
 
Sheet Metal Forming Test Adapter
Rolling Four Ball
Small Thrust Washer
Single Ball-on-Disk
 
Standard Multi specimen adapters can be used up to a temperature of 230¡ÆC.
 
   E-mail: lab@labkorea.com
ESEL TechTra Inc.
 #1382-4, Kuweol-3dong, Namdongku, Incheon, korea
 ©Copyright 1998 www.labkorea.com ® All right reserved