Large steam sterilizer
   HS-700S, HS-1000S, HS-1000D
Middle steam sterilizer
   
HS-4085S, HS-5020S, HS-5025S/D
   HS-5035S/D, HS-250, HS-350S
Semi-automatic steam sterilizer
   HS-85, HS-196
Vacuumclave steam sterilizer
   HS-2321V, HS-3041V
Autoclave
  - Gravity type steam sterilizer
    HS-1212, HS-1321, HS-9041
  -  Vertical type steam sterilizer
    HS-60
Field hospital steam sterilizer
   HS-1636
EO gas sterilizer
   | HS-4085EO | | HS-700EO |
Ultraviolet rays sterilizer
   H-3000B, H-5000A, H-8000C

 

Large steam sterilizer

 | specification | | cycle graph |

 

HS-700S

HS-1000S

HS-1000D

 | Feature |

 1. CONTROL SYSTEM

The Control System with a built-in microprocessor automatically controls all programs including sterilization cycle, digital display, self-diagnosis and printing.

- While in operation, it measures, controls and shows in digital display the time, temperature and pressure, etc.
- While the power is failed, the Real Time Clock driven by its own back-up battery keeps displaying the present year, month, and time.
- If the power supply is failed while in cycle, it maintains the present state by its own bank-up battery; If the power supply is resumed within ine minute, continue the cycle which was in execution and of one minute has passed, the cycle is disposed to an error.
- The Control System is of a highly compact type; with one-touch of starter the chamber door is locked and then all phases such as condition, sterilization, exhaust, drying, completion, printing and even the door unlocking are automatically performed.

2. CYCLE PROGRAM

It has the built -in program consisting of 5 kinds of basic cycles and 2 kinds of test cycles which can changed at the option of users if it is necessary. It is noteworthy that the liquid cycle is supposed to have a 30-minute cooling time to ensure the safety of sterilization

3. SELF - DIAGNOSIS

While in operation, if the follwing troubles occurs, it senses this and displays ¡¸warning code¡¹ or ¡¸caution code¡¹ in the message sound, aborting of the cycle etc.

4. PRINTER

It prints out two kinds of the cycle data for interpreting and keeping . One is the ¡¸Temperature Graph¡¹ showing the changing pattern temperature of the cycle and other is the ¡¸Cycle Performance Record¡¹printing the quantifled change of the time, temperature, pressure and summarized data of the cycle.

5. CONTROL UNIT

The Control System do important functions as the brain of sterilizer; if any failure occurs, it causes an immense trouble to the hospital work unless the prompt actions are taken. The controls of this sterilizer os of the compact unit type which any person can easily replacement of Control Unit can be supplied promptly by the express delivery service, thereby minimizing the inconvenience to the user.

¡á FEATURES OF DOUBLE DOOR TYPE

The sterilizer chamber is equipped with a door on both sides; a wall is installed on both sides of the chamber in the center to shut off the space divided into loading area and unloading area. This is designed to shut off the air circulation between the space of both sides in order to prevent the sterilized items from being contaminated again. To this purpose, both doors do not open at the same time, but only one door will open; the loading of goods and the controlling of cycle takes place in the loading area and the sterilized goods are taken out in the unloading area.

¡á specification

MODEL

HS-700S

HS-1000S

HS-1000D

Type

General Purpose, Steam Powered

General Purpose, Steam Powered

General Purpose, Steam Powered

Style

For Recessing, Single Door

For Recessing, Single Door

For Recessing, Double Door

Overall Size

W1,160×H1,881×D1,652mm

W1,300×H1,894×D1,705mm

W1,300×H1,894×D2,036mm

Chamber 
Size & Cap

W610×H914×D1,286mm

W680×H1,200×D1,286mm

W680×H1,200×D1,316mm

717§¤ ,(SUS316)

1,049§¤ ,(SUS316)

1,073§¤ ,(SUS316)

Door

Manual Operation

Power/Manual Operation

Power/Manual Operation

Operating
Temperature

100¡É¡­135¡É

100¡É¡­135¡É

100¡É¡­135¡É

Air Removal

Gravity / Vacuum

Gravity / Vacuum

Gravity / Vacuum

Vacuum System

Water Pump : 220V/530W,

Water Pump : 220V/800W

Water Pump : 220V/800W

Water Ejector

Water Ejector

Water Ejector

Steam Source

Saturated Steam of min.
97% Dryness

Saturated Steam of min.
97% Dryness

Saturated Steam of min.
97% Dryness

Pressure : 3.5¡­5kgf/cm2

Pressure : 3.5¡­5kgf/cm2

Pressure : 3.5¡­5kgf/cm2

Flow Rate : max.150kg/hr,
Avg. 60kg/hr

Flow Rate : max.200kg/hr,
Avg. 70kg/hr

Flow Rate : max.200kg/hr,
Avg. 70kg/hr

Connection on Unit : 3/4"FNPT

Connection on Unit : 3/4"FNPT

Connection on Unit : 3/4"FNPT

Water Source

Clean Water of 20¡Émax

Clean Water of 20¡Émax

Clean Water of 20¡Émax

Pressure : 2.1¡­5kgf/cm2

Pressure : 2.1¡­5kgf/cm2

Pressure : 2.1¡­5kgf/cm2

Flow Rate : max.50§¤/min
Avg.1,200kg/hr

Flow Rate : max.60§¤/min
Avg.2,000kg/hr

Flow Rate : max.60§¤/min
Avg.2,000kg/hr

Connection on Unit : 3/4"FNPT

Connection on Unit : 1"FNPT

Connection on Unit : 1"FNPT

Drain

Connection on Unit: 1"FNPT

Connection on Unit: 1¡­1/4" FNPT

Connection on Unit: 1¡­1/4" FNPT

Exclusive Drain each Unit: 2"metallic pipe

Exclusive Drain each Unit: 2"metallic pipe

Exclusive Drain each Unit: 2"metallic pipe

Power Source

AC220V,1ph,50/60hz,15Amps

AC220V,1ph,50/60hz,15Amps

AC220V,1ph,50/60hz,15Amps

Control

Microprocessor

Microprocessor

Microprocessor

Display

LCD with EL backilght, 240x64Dots,Graphic Display

Printer

Dot Matrix,42 Columns, Single Color,Paper Roll: W76 x Ø80mm

Loading System

Loading car : 1ea

Loading car : 1ea

Loading car : 1ea

Carriage : 1ea

Carriage : 1ea

Carriage : 2ea

Net weight

Total : 1,420kg

Total : 1,767kg

Total : 2,390kg

Unit : 1,306kg

Unit : 1,656kg

Unit : 2,279kg

Loading Car : 30kg

Loading Car : 42kg

Loading Car : 42kg

Carriage : 84kg

Carriage : 69kg

Carriage : 69kg

Operational
Environment
Conditions
(IEC 601-1)

¨ÍAn ambient temperature range of +10¡É to+40¡É.

¨ÎA relative humidity range of 30% to 75%.

¨ÏAn atmospheric pressure range of 700hPato 1,060hPa.

¨ÐA temperature of the water at the inlet of water -cooled EQUIPMENT not higher than 25¡É.

 

¡á cycle graph

 ¡å CYCLE PROGRAM


PRESSURE

TEMPERATURE




CYCLE NO.1 PRE-VACUUM WRAPPED

 

             
CYCLE NO.2 GRAVITY WRAPPED      CYCLE NO.3 GRAVITY UNWRAPPED  CYCLE NO.4 GRAVITY UNWRAPPED
             
CYCLE NO.5 GRAVITY LIQUID                    VACUUM LEAK TEST                        BOWIE-DICK TEST

 

 ¡å STERILIZATION CYCLE PROGRAM

 

CYCLE No.

1

2

3

4

5

Air Removal

Vacuum

Gravity

Gravity

Gravity

Gravity

Loads Type

Wrapped

Wrapped

Unwrapped

Unwrapped

Liquid

Exposure Temperature

132¡É

132¡É

132¡É

121¡É

121¡É

Exposure Time

5 min

15 min

15 min

30 min

30 min

Dry Time

5 min

15 min

0

0

0

Cooling Time

0

0

0

0

30 min

 

¡å TEST CYCLE PROGRAM

 

ITEM

Vacuum Leak Test

Bowie-Dick Test

Cycle No.

1

2

Test Pack

N/A

Specific Pack

Process

Operationg Timp.: 132¡É
Exposure Time: 3 min
Dry Time: 15 min
Dwell Time: 5 min
Test Time: 15 min

Operationg Temp. : 134¡É
Exposure Time
: 3.5 min