Large steam sterilizer | specification | | cycle graph | |
|
|
|
|
|||||
| Feature | 1. CONTROL SYSTEM The Control System with a built-in
microprocessor automatically controls all programs including
sterilization cycle, digital display, self-diagnosis and printing. - While in operation, it measures,
controls and shows in digital display the time, temperature
and pressure, etc. |
|||||
2. CYCLE PROGRAM It has the built -in program consisting of 5 kinds of basic cycles and 2 kinds of test cycles which can changed at the option of users if it is necessary. It is noteworthy that the liquid cycle is supposed to have a 30-minute cooling time to ensure the safety of sterilization 3. SELF - DIAGNOSIS While in operation, if the follwing troubles occurs, it senses this and displays ¡¸warning code¡¹ or ¡¸caution code¡¹ in the message sound, aborting of the cycle etc. 4. PRINTER It prints out two kinds of the cycle data for interpreting and keeping . One is the ¡¸Temperature Graph¡¹ showing the changing pattern temperature of the cycle and other is the ¡¸Cycle Performance Record¡¹printing the quantifled change of the time, temperature, pressure and summarized data of the cycle. 5. CONTROL UNIT The Control System do important functions as the brain of sterilizer; if any failure occurs, it causes an immense trouble to the hospital work unless the prompt actions are taken. The controls of this sterilizer os of the compact unit type which any person can easily replacement of Control Unit can be supplied promptly by the express delivery service, thereby minimizing the inconvenience to the user. ¡á FEATURES OF DOUBLE DOOR TYPE The sterilizer chamber is equipped with a door on both sides; a wall is installed on both sides of the chamber in the center to shut off the space divided into loading area and unloading area. This is designed to shut off the air circulation between the space of both sides in order to prevent the sterilized items from being contaminated again. To this purpose, both doors do not open at the same time, but only one door will open; the loading of goods and the controlling of cycle takes place in the loading area and the sterilized goods are taken out in the unloading area. |
MODEL |
HS-700S |
HS-1000S |
HS-1000D |
Type |
General Purpose, Steam Powered |
General Purpose, Steam Powered |
General Purpose, Steam Powered |
Style |
For Recessing, Single Door |
For Recessing, Single Door |
For Recessing, Double Door |
Overall Size |
W1,160×H1,881×D1,652mm |
W1,300×H1,894×D1,705mm |
W1,300×H1,894×D2,036mm |
Chamber |
W610×H914×D1,286mm |
W680×H1,200×D1,286mm |
W680×H1,200×D1,316mm |
717§¤ ,(SUS316) |
1,049§¤ ,(SUS316) |
1,073§¤ ,(SUS316) |
|
Door |
Manual Operation |
Power/Manual Operation |
Power/Manual Operation |
Operating |
100¡É¡135¡É |
100¡É¡135¡É |
100¡É¡135¡É |
Air Removal |
Gravity / Vacuum |
Gravity / Vacuum |
Gravity / Vacuum |
Vacuum System |
Water Pump : 220V/530W, |
Water Pump : 220V/800W |
Water Pump : 220V/800W |
Water Ejector |
Water Ejector |
Water Ejector |
|
Steam Source |
Saturated Steam of min. |
Saturated Steam of min. |
Saturated Steam of min. |
Pressure : 3.5¡5kgf/cm2 |
Pressure : 3.5¡5kgf/cm2 |
Pressure : 3.5¡5kgf/cm2 |
|
Flow Rate : max.150kg/hr, |
Flow Rate : max.200kg/hr, |
Flow Rate : max.200kg/hr, |
|
Connection on Unit : 3/4"FNPT |
Connection on Unit : 3/4"FNPT |
Connection on Unit : 3/4"FNPT |
|
Water Source |
Clean Water of 20¡Émax |
Clean Water of 20¡Émax |
Clean Water of 20¡Émax |
Pressure : 2.1¡5kgf/cm2 |
Pressure : 2.1¡5kgf/cm2 |
Pressure : 2.1¡5kgf/cm2 |
|
Flow Rate : max.50§¤/min |
Flow Rate : max.60§¤/min |
Flow Rate :
max.60§¤/min |
|
Connection on Unit : 3/4"FNPT |
Connection on Unit : 1"FNPT |
Connection on Unit : 1"FNPT |
|
Drain |
Connection on Unit: 1"FNPT |
Connection on Unit: 1¡1/4" FNPT |
Connection on Unit: 1¡1/4" FNPT |
Exclusive Drain each Unit: 2"metallic pipe |
Exclusive Drain each Unit: 2"metallic pipe |
Exclusive Drain each Unit: 2"metallic pipe |
|
Power Source |
AC220V,1ph,50/60hz,15Amps |
AC220V,1ph,50/60hz,15Amps |
AC220V,1ph,50/60hz,15Amps |
Control |
Microprocessor |
Microprocessor |
Microprocessor |
Display |
LCD with EL backilght, 240x64Dots,Graphic Display |
||
Printer |
Dot Matrix,42 Columns, Single Color,Paper Roll: W76 x Ø80mm |
||
Loading System |
Loading car : 1ea |
Loading car : 1ea |
Loading car : 1ea |
Carriage : 1ea |
Carriage : 1ea |
Carriage : 2ea |
|
Net weight |
Total : 1,420kg |
Total : 1,767kg |
Total : 2,390kg |
Unit : 1,306kg |
Unit : 1,656kg |
Unit : 2,279kg |
|
Loading Car : 30kg |
Loading Car : 42kg |
Loading Car : 42kg |
|
Carriage : 84kg |
Carriage : 69kg |
Carriage : 69kg |
|
Operational |
¨ÍAn ambient temperature range of +10¡É to+40¡É. |
||
¨ÎA relative humidity range of 30% to 75%. |
|||
¨ÏAn atmospheric pressure range of 700hPato 1,060hPa. |
|||
¨ÐA temperature of the water at the inlet of water -cooled EQUIPMENT not higher than 25¡É. |
|||
|
¡å CYCLE PROGRAM
|
|
¡å STERILIZATION CYCLE PROGRAM
CYCLE No. |
1 |
2 |
3 |
4 |
5 |
Air Removal |
Vacuum |
Gravity |
Gravity |
Gravity |
Gravity |
Loads Type |
Wrapped |
Wrapped |
Unwrapped |
Unwrapped |
Liquid |
Exposure Temperature |
132¡É |
132¡É |
132¡É |
121¡É |
121¡É |
Exposure Time |
5 min |
15 min |
15 min |
30 min |
30 min |
Dry Time |
5 min |
15 min |
0 |
0 |
0 |
Cooling Time |
0 |
0 |
0 |
0 |
30 min |
¡å TEST CYCLE PROGRAM
ITEM |
Vacuum Leak Test |
Bowie-Dick Test |
Cycle No. |
1 |
2 |
Test Pack |
N/A |
Specific Pack |
Process |
Operationg Timp.: 132¡É |
Operationg Temp. : 134¡É |
|