Middle steam sterilizer | Feature | | Specification | |
|
|
|
|
|
||||||
| Feature | | Specification | |
|
|
||||
|
||||||
The
sterilizer is equipped with a practical size of rectangular
or cylindrical sterilizing chamber. The sterilizing process
is most efficient because the materials are sterilized and dreed
by air femobal with the pre-vacuum and post vacuum. In particular,
the sterilizer with a double door style prevents recontamination
by retrieving processed materials to the clean room. 1. CONTROLS The Control System with a built- in microprocessor, controls process of the cycle, the function of display of a large 5"LCD showing the operating conditions of the sterilizer, the self-diagnosis which detects abnormality and takes necessaly actions and printing which prints out cycle performance records automatically. |
||||||
2. CYCLE PROGRAM - The Basic Cycles are programs generally used
for most materials to be sterilized. 3. BOWIE-DICK TEST PROGRAM This program provides a test for air residual in the chamber after air removal by the vacuum system. If this program detects any residual of air, the wrapped materials shall mot be processed with the vacuum cycle program. 4. LEAK TEST PROGRAM The leak test program, designed for test of any leakage out of the chamber, shall be performed prior to vacuum sterilization cycle once everyday. If any leakage is detected throughout the test, the vacuum sterilization cycle would not operate. 5. AIR FILTER The bacteria-retentive filter which can filter out 99.97% of 0.3§ particles filters outside air supplied into the chamber during the vent and drying the vent and drying phase. If prevents sterilized items from recontamination by bacteria in the air. 6. VACUUM SYSTEM The vacuum system with the water pump and ejector is an equipment used for exhausting air or vapor out of the chamber. If enhances sterilization and drying effect with a strong vacuum force and minimizes noise and malfunction. 7. DOOR INTERLOCK For safety, the chamber door would not open when the power is turned off, the cycle would not operate when the door is open and the door would not open when any pressure exists in the chamber. In a case of the double door style, both of doors would not open simultaneously and one of the doors would open only the other one is locked. 8. SELF-DIAGNOSIS Seif-diagnosis, detecting 12 types of error, generates alarm and automatically takes any necessary actions. 9. PRINTER It prints a cycle performance record which reviews, analyzes and saves sterilization results related to management. This document includes sterilizer number, date and yeat, setting value(sterilization temperature, condition temperature, exposure time, and drying time), performance data of each phase (start and finish time, temperature and pressure), maximum/minimum temperature and pressure, elapse time of each phases, acceptance or unacceptance of the cycle and other data (cycle time, cycle number, error codes, power failure/recovery time). |
| Specification |
MODEL |
HS-4085S |
HS-5020S |
Cabinet |
W800×H 1,600×D 892mm |
W900×H 1,600×D 1,122mm |
Chamber |
Stainless Steel(SUS 316) |
Stainless Steel(SUS 316) |
Cylindrical Type w/Jacker |
Cylindrical Type w/Jacker |
|
85.4§¤cap., ø400×D680(mm) |
196§¤cap., ø500×D1000(mm) |
|
Operation |
100¡É¡135¡É |
100¡É¡135¡É |
Air Removal |
Vacuum/Gravity |
Vacuum/Gravity |
Power Source |
Standard (Built-in Steam Generator and External Steam System): |
Standard (Built-in Steam Generator and External Steam System): |
AC220V, 3ph, 3wire,50/60§Ô, 26A or |
AC220V, 3ph, 3wire,50/60§Ô, 32A or |
|
AC380V, 3ph, 4wire,50/60§Ô, 15A |
AC380V, 3ph, 4wire,50/60§Ô, 20A |
|
Optional(External
Steam System): |
Optional(External
Steam System): |
|
Controls |
8bit Microprocessor |
8bit Microprocessor |
Steam Source |
Steam
Generator :Heater 9§Ò, |
Steam
Generator :Heater 12§Ò, |
External
Steam System : |
External
Steam System : |
|
Pressure
: 3.5§¸f/§²¡5§¸f/§², |
Pressure
: 3.5§¸f/§²¡5§¸f/§², |
|
Water Source |
Pressure:2§¸f/§²¡5§¸f/§², |
Pressure:2§¸f/§²¡5§¸f/§², |
Vacuum System |
Water
Pump : 220V / 530W, |
Water
Pump : 220V / 530W, |
Loading System |
Shelves : W300 × L360(mm),2 set, |
Shelves : W450 × L450(mm),2 set, |
Rack : 1 set |
Rack : 1 set |
|
Net Weight |
180kg |
270kg |
Display |
160×128 Dots Graphic Display, LCD (EL back-light) |
|
printer |
Dot Matrix, 40 Columns, Single Color Paper roll: W57.5mm×L22m |
|
Operational |
¨ÍAn ambient temperature range of + 10¡É
to + 40¡É. |
|
MODEL |
HS-5025S |
HS-5025D |
HS-5035S |
HS-5035D |
Chbinet |
W1,000×H1,740×D1,172mm |
W1,000×H1,740×D1,109mm |
W1,000×H1,740×D1,172mm |
W1,000×H1,740×D1,109mm |
Chamber |
Stainless Steel(SUS 316) |
Stainless Steel(SUS 316) |
Stainless Steel(SUS 316) |
Stainless Steel(SUS 316) |
Rectangular Type |
Rectangular Type |
Rectangular Type |
Rectangular Type |
|
250§¤cap. |
250§¤cap. W500×H500×D1,000mm |
350§¤cap. W500×H700×D1,000mm |
350§¤cap. W500×H700×D1,000mm |
|
Single Door |
Double Door |
Single Door |
Double Door |
|
Operation |
100¡É¡135¡É |
100¡É¡135¡É |
100¡É¡135¡É |
100¡É¡135¡É |
Air Removal |
Vacuum/Gravity |
Vacuum/Gravity |
Vacuum/Gravity |
Vacuum/Gravity |
Power Source |
Standard (Built-in Steam Generator and External Steam System): |
|||
AC 220V, 3ph, 50/60§Ô, 55A or AC 380V, 3ph, 4wire, 50/60§Ô, 32A |
||||
Optional(External Steam System):AC220V, 50/60§Ô, 5A |
||||
Controls |
8 bit Microprocessor |
8 bit Microprocessor |
8 bit Microprocessor |
8 bit Microprocessor |
Steam Source |
Steam Generator: |
Steam Generator: |
Steam Generator: |
Steam Generator: |
Water Pump : 300W |
Water Pump : 300W |
Water Pump : 300W |
Water Pump : 300W |
|
External Steam System: |
External Steam System: |
External Steam System: |
External Steam System: |
|
Saturated Steam of 97% Dryness |
Saturated Steam of 97% Dryness |
Saturated Steam of 97% Dryness |
Saturated Steam of 97% Dryness |
|
Pressure
: |
Pressure
: |
Pressure
: |
Pressure
: |
|
Flow Rate: max.30§¸/hr |
Flow Rate: max.30§¸/hr |
Flow Rate: max.30§¸/hr |
Flow Rate: max.30§¸/hr |
|
Water Source |
Pressure: |
Pressure: |
Pressure: |
Pressure: |
Flow Rate : max.50§¤/min |
Flow Rate : max.50§¤/min |
Flow Rate : max.50§¤/min |
Flow Rate : max.50§¤/min |
|
Vacuum System |
Water Pump
: |
Water Pump
: |
Water Pump
: |
Water Pump
: |
Water Ejector |
Water Ejector |
Water Ejector |
Water Ejector |
|
Loding System |
Shelves : |
Shelves : |
Shelves : |
Shelves : |
Rack : 1set |
Rack : 1set |
Rack : 1set |
Rack : 1set |
|
Net Weight |
665kg |
755kg |
770kg |
890kg |
Display |
160×128 Dots Graphic Display, LCD (EL back-light) |
|||
Printer |
Dot Matrix, 40 Columns, Single Color Paper roll: W57.5mm×L22m |
|||
Operational |
¨ÍAn ambient temperature range of + 10¡É
to + 40¡É. |
|||
|
Feature | |
1. The Temperature Controller/Recorder precisely controls the set temperature for sterilization and at the same time, the sterilization temperature is recorded in graph on the chart for 24 hours whereby the result of sterilization can be checked; it can also serve as the thermometer indicating the present temperature. |
2. It is equipped with the vacuum system consisting of powerful water pump and ejector, therefore it performs the air removal as the prevacuum and performs drying phase as the post vacuum after sterilization phase. consequently, it effective in improving the penetration of sterilization agent and the drying performance, and shortening the cycle time and reducing the noise and the trouble rate of equipment. |
3. If the door is not locked, the cycle is not started; if the pressure of 0.3§¸f/§² or more remains in the chamber after sterilization, the safety device prevents the door from being opened. |
4. The standard type HS-250 uses the built-in steam generator exclusively for the steam source while the standard type HS-350s can use exclusively the external steam. But either the built -in steam generator or the external supply system can be exclusively used; otherwise both of them can be concurrently used regardless model type upon the request of the user. |
5. The built-in steam generator is equipped with electric heater,automatic temperature regulator, automatic water feed system and safety valves, etc.; it is the device which ensures a stable supply of saturated steam whereby the sterilizer (HS-250) can be used without the limitation of location. |
| Specification |
MODEL |
HS-250 |
HS-350S |
Cabinet |
Stainless Steel Plate |
Stainless Steel Plate |
W765×H1,855×D1,168mm |
W765×H1,855×D1,168mm |
|
Chamber |
Rectangular w/Jacket(SUS316) |
Rectangular w/Jacket(SUS316) |
W500×H500×D1,000mm,250§¤cap |
W500×H700×D1,000mm,350§¤cap |
|
Air Removal |
Gravity / Vacuum |
Gravity / Vacuum |
Operating |
121¡É(1.0§¸f/§²)¡132¡É(2.0§¸f/§²) |
121¡É(1.0§¸f/§²)¡132¡É(2.0§¸f/§²) |
Power Source |
Standard : Built-in Steam Generator |
Standard : External Steam Generator |
AC220V,3ph,50/60§Ô,47A or |
AC220V,1ph,50/60§Ô,5A |
|
AC380V,3ph,4wire,50/60§Ô,31A |
¡¡ |
|
Optional : External Steam System |
Optional : Built-in Steam Generator |
|
AC220V,1ph,50/60§Ò,5A |
AC220V,3ph,50/60§Ô,47A or |
|
¡¡ |
AC380V,3ph,4wire,50/60§Ô,31A |
|
Steam Source |
Standard : Built-in Steam Generator |
Standard : External Steam System |
1,5bhp, max. 7§¸f/§² |
Saturated Steam of min. 97% Dryness |
|
Heater : 15§Ò |
Pressure: 3.5§¸f/§²¡5§¸f/§² |
|
Feed Pump : 250W |
Flow Rate: max.50§¸/hr |
|
Optional: External Steam System |
Optional : Built-in Steam Generator |
|
Saturated Steam of min. 97% Dryness |
1,5bhp, max. 7§¸f/§² |
|
Pressure: 3.5§¸f/§²¡5§¸f/§² |
Heater : 15§Ò |
|
Flow Rate: max.50§¸/hr |
Feed Pump : 250W |
|
Water Source |
Pressure : 2§¸f/§²¡5§¸f/§² |
Pressure : 2§¸f/§²¡5§¸f/§² |
Flow Rate : max.50§¤/min |
Flow Rate : max.50§¤/min |
|
Vacuum System |
Water Pump : 880W, Water Ejector |
Water Pump : 880W, Water Ejector |
Control System |
Temperature Indicator/Controller/Recorder |
Temperature Indicator/Controller/Recorder |
Adjustable Range : 70¡É¡130¡É |
Adjustable Range : 70¡É¡130¡É |
|
Chart : Ø6", 1rpd |
Chart : Ø6", 1rpd |
|
Timer |
Electric Powered, Auto-Reset Type 0¡60min |
Electric Powered, Auto-Reset Type 0¡60min |
for Sterilizing & Drying |
for Sterilizing & Drying |
|
Loading System |
Shelves : W450×L906mm,3ea |
Shelves : W450×L906mm,4ea |
Rack : 1ea |
Rack : 1ea |
|
Net Weight |
674§¸ |
650§¸ |
Operational |
¨ÍAn ambient temperature range of +10¡É to
+40¡É. |
|
|