UV/VIS SPECTROPHOTOMETER OPTIZEN 3220UV

SPECIFICATION

Spectral Band Width

1<nm

Wavelength Range

190nm-1100nm

Wavelength Accuracy

± 0.3nm

Wavelength Repeatability

± 0.1 nm

Monochromater

Modified Czemy -Turner type with blazed Grating (lb = 313.2 or 253.7)

Photometric system

Double Beam Optics

Monochromator

Modified Czerny-Turner type with 1200 lines/mm blazed Grating

Scan Speed

300nm/min ~ 3000nm/min

Stray Light

< 0.05%T at 220nm and 340nm

Photometric display range

-0.5 to 4.0A

Photometric accuracy

± 0.004A at 1A

Reproducibility

± 0.002A

Noise level

<0.001A around 0A at 220 and 340nm

Drift 0.002A/hr at 340nm
Stability < ± 0.002 A/h
Light Source Tungsten-Halogen Lamp & Deuterium Lamp
Display Mode 17″ TFT LCD Monitor
Sample holder Automatic rotary type 8-position multi-cell holder
Wavelength Slew rate 5000 nm/min
Standard capability Absorbance/% Transmittance, Concentration and file manager
Power Requirement 220V/1A, 60Hz

 

Software Specifications

1. ATC + STC

Display of all data - %T , Abs, Concentration

Measure absorbance at up to 10 wavelengths

Save Methods up to 999 each
Selected unit
Single-point calibration curve (zero or Non-zero) , or multipoint calibration curve method (1st ~ 3rd order regression curves or segmented standard curve).
Manual standard curve : input concentration and Abs up to 30
Save data in Excel form
Measure different wavelengths for each cell holder
Manually delete the wrong datum when first make calibration curve
 

2. Survey Scan

Scan resolution : up to 0.1nm

Save methods, up to 999 each

Scale mode-manual, auto, cursor, peaks valleys
Abs. <--> %T conversion
Graph manipulation mode
Make Calibration curve by multi scaning data
Curve fitting mode
Save data in Excel & BMP form
Peak/Valley auto detection
 

3. KINETIC

Measurement time : total time: 3hr, interval time : ≥1 sec

Data view type : graph or table

Scale mode-manual, auto, cursor, peaks valleys
Abs. <--> %T conversion
Graph manipulation mode
Peak/Valley auto detection
Save data in Excel & BMP form
Save methods, up to 999 each